ISSN 2063-5346
For urgent queries please contact : +918130348310

Analytical method development and validation of RP-HPLC method for quantification of Sorafenib tosylate loaded solid Lipid Nanoparticles

Main Article Content

Uttam kumar, R.S Chandan, Srividaya Gorantla, Hemanth Vikram, Sudip pal, Abhishek kumar Dev, Muskan, Anand kumar Tengli
» doi: 10.31838/ecb/2023.12.si4.088

Abstract

This study set out to create an RP-HPLC system that was effective, sensitive, picky, precise,accurate and practical. For this, a UV detection technique for detecting Sorafenib tosylate-loaded solid lipid nanoparticles has been developed and validated. To improve the procedure, many parameters were used (pH and Column). The chromatographic separation was carried out using a Shimadzu prominence-i LC-2030C and a C8 short column (5 m 4.6 x 100 mm). With a 10 minute runtime, a 10 mL injection volume, and a 1 mL/min flow rate, the mobile phase is a mixture of 70:30 methanol:0.1% formic acid in water. The effluent is detected at 261nm using a UV detector. Drug Entrapment Efficiency (DEE) and Drug Loading (DL) for ST from the extracted SLNs matrix were found to be 86.9% and 19%, respectively. The developed analytical method has a linearity range of 1-64g/ml and an R2 value of 0.998. 0.88 g/ml detection limit (LOD) and 1.0 g/ml limit of quantification (LOQ), and 0.88 g/ml detection limit (LOQ). Using ICH Q2(R1) guidelines, the proposed technique was evaluated, and it was shown to be accurate, linear, robust, and specific. Using the devised analytical method, drug release, drug loading, and drug entrapment effectiveness were all studied.

Article Details